Avaliação do comportamento mecânico de prótese fixa implantossuportada com cantilever por meio da técnica de extensometria elétrica
Devanir Silva Vieira Prado
TESE
Português
T/UNICAMP P882a
[Evaluation of mechanical behavior of fixed prosthesis with cantilever implant through technical electrical extensometry]
Piracicaba, SP : [s.n.], 2013.
83 f. : il.
Orientador: Wilkens Aurelio Buarque e Silva
Tese (doutorado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba
Resumo: O presente trabalho avaliou o comportamento mecânico de próteses fixas implantossuportadas submetidas às cargas oblíquas com 20 graus de inclinação. Foi confeccionado um corpo-de-prova em liga metálica à base de CoCr, simulando uma prótese fixa com dois cantiléveres, suportada por seis...
Ver mais
Resumo: O presente trabalho avaliou o comportamento mecânico de próteses fixas implantossuportadas submetidas às cargas oblíquas com 20 graus de inclinação. Foi confeccionado um corpo-de-prova em liga metálica à base de CoCr, simulando uma prótese fixa com dois cantiléveres, suportada por seis implantes separados pela distância de 10mm. Foram colados quatro extensômetros em cada um dos pilares (UCLA), um em cada face (mesial, distal, vestibular e lingual) e dois em cada cantilever. Após posicionamento da infra-estrutura e o torque nos parafusos, o conjunto foi submetido ao ensaio mecânico com carga compressiva oblíqua de 20 graus, com 15 Kgf na região do cantilever direito, a 7mm do pilar distal. A velocidade de aplicação de carga foi de 15 Kgf por segundo, com frequência de 0,5Hz totalizando 5 ciclos de aplicação. Foram avaliadas as deformações ocorridas em cada extensômetro de um mesmo pilar e em todos os pilares do corpo-de-prova. A média dos valores obtidos foi analisada pelos testes ANOVA e Tukey-Kramer, com significância de 5%. Diferenças significantes foram encontradas entre as faces de um mesmo pilar e entre as faces de todos os pilares avaliados (p<0,0001). Com aplicação da carga oblíqua no cantilever, o pilar adjacente foi o mais solicitado por compressão e os pilares mais anteriores da mesma forma em menor intensidade. Foi observado que os pilares anteriores foram submetidos ao regime de tração (alavanca)
Ver menos
Abstract: This study aimed to evaluate the mechanical behavior of implant-supported fixed prostheses, with oblique loads submitted to the 20° tilt. Was made in a body of evidence-based alloy CoCr, simulating a fixed prosthesis with two cantilever supported by six implants separated by a distance of...
Ver mais
Abstract: This study aimed to evaluate the mechanical behavior of implant-supported fixed prostheses, with oblique loads submitted to the 20° tilt. Was made in a body of evidence-based alloy CoCr, simulating a fixed prosthesis with two cantilever supported by six implants separated by a distance of 10mm. Four strain gauges were glued in each of the pillars (UCLA), one on each face (mesial, distal, buccal and lingual) and two on each cantilever. After the torque of the pillars and the infrastructure, the set was subjected to a mechanical test with oblique compressive loads of 20°, with 15Kgf in the region of the cantilever law, 7mm distal abutment. The rate of application of load was 15N per second at a frequency of 5 cycles totaling 0.5 HTZ applications. We evaluated the deformations occurring in each extensometer of one pillar in all pillars body of evidence. The mean values were analyzed by ANOVA and Tukey-Kramer, with a significance of 0.05. Significant differences were found between the faces of the same pillar and between the faces of all the pillars assessed (p <0.0001). With application of the cantilever load oblique, the adjacent pillar was the most requested by compressing and the pillars likewise earlier in smaller amounts. It was observed that the pillars were requested earlier also under traction (lever)
Ver menos
Silva, Wilkens Aurelio Buarque e, 1967-
Orientador
Oliveira, Milene de, 1981-
Avaliador
Hermann, Caio
Avaliador
Landulpho, Alexandre Brait
Avaliador
Ramos, Guilherme da Gama, 1976-
Avaliador
Avaliação do comportamento mecânico de prótese fixa implantossuportada com cantilever por meio da técnica de extensometria elétrica
Devanir Silva Vieira Prado
Avaliação do comportamento mecânico de prótese fixa implantossuportada com cantilever por meio da técnica de extensometria elétrica
Devanir Silva Vieira Prado
Exemplares
Nº de exemplares: 1
Não existem reservas para esta obra