Stress evaluation of implant-abutment connections under different loading conditions : a 3D finite element study
Jessica Mie Ferreira Koyama Takahashi, Andreza Costa Dayrell, Rafael Leonardo Xediek Consani, Mauro Antônio de Arruda Nóbilo, Guilherme Elias Pessanha Henriques, Marcelo Ferraz Mesquita
ARTIGO
Inglês
This study evaluated the effects of axial and oblique occlusal loading on implant-supported partial dentures with different connection systems (external hexagon, internal hexagon, and Morse taper). Upon axial loading, all systems presented similar stress values. Stress values increased under oblique...
Ver mais
This study evaluated the effects of axial and oblique occlusal loading on implant-supported partial dentures with different connection systems (external hexagon, internal hexagon, and Morse taper). Upon axial loading, all systems presented similar stress values. Stress values increased under oblique loading. Stress distribution changed for some of the internal connection structures. It can be concluded that oblique load increases stress on bone structures and prosthetic components. Internal connection system implants present more favorable stress distribution patterns than do external connection system implants.
Ver menos
COORDENAÇÃO DE APERFEIÇOAMENTO DE PESSOAL DE NÍVEL SUPERIOR - CAPES
Fechado
Stress evaluation of implant-abutment connections under different loading conditions : a 3D finite element study
Jessica Mie Ferreira Koyama Takahashi, Andreza Costa Dayrell, Rafael Leonardo Xediek Consani, Mauro Antônio de Arruda Nóbilo, Guilherme Elias Pessanha Henriques, Marcelo Ferraz Mesquita
Stress evaluation of implant-abutment connections under different loading conditions : a 3D finite element study
Jessica Mie Ferreira Koyama Takahashi, Andreza Costa Dayrell, Rafael Leonardo Xediek Consani, Mauro Antônio de Arruda Nóbilo, Guilherme Elias Pessanha Henriques, Marcelo Ferraz Mesquita
Fontes
Journal of oral implantology (Fonte avulsa) |