Title: Magnitude de artefatos de tomografia computadorizada de feixe cônico relacionados com diferentes tipos de implantes dentários : impacto na qualidade da imagem = Magnitude of CBCT image artifacts related to dental implants: impact on image quality
Title Alternative: Magnitude of CBCT image artifacts related to dental implants : impact on image quality
Author: Fontenele, Rocharles Cavalcante, 1995-
Advisor: Freitas, Deborah Queiroz de, 1977-
Abstract: Resumo: A presença de artefatos em imagens de Tomografia Computadorizada de Feixe Cônico (TCFC) é inevitável, quando o feixe de radiação interage com materiais de alto número atômico, como implantes dentários. De acordo com a literatura consultada, ainda não existem estudos que tenham avaliado a magnitude desse efeito para os diferentes materiais de implantes dentários. Assim, o objetivo no presente estudo foi avaliar a magnitude dos artefatos produzidos pela presença de implantes de titânio e de zircônia em imagens de TCFC. Para isso, duas mandíbulas humanas foram utilizadas para a inserção de um implante de titânio e um de zircônia, em diferentes momentos, na região do dente 46. Além disso, na cortical vestibular, na região do implante dentário, foi fixado um bloco com densidade semelhante à do osso com a finalidade de padronizar a seleção das imagens a serem avaliadas. Cada mandíbula humana foi mantida, com o auxílio de um material de moldagem, em um container cilíndrico de plástico (16 cm de diâmetro) preenchido com água para simulação da atenuação do feixe de radiação pelos tecidos moles. As aquisições das imagens foram realizadas em três aparelhos de TCFC: Picasso Trio 3D, ProMax 3D e 3D Accuitomo 80, utilizando o mesmo protocolo de aquisição: 80 kVp, 5 mA e voxel de tamanho 0,2 mm. Os exames foram adquiridos antes e após a inserção dos implantes. Para a avaliação das imagens, uma mesma reconstrução axial foi selecionada, utilizando como referência o bloco, na região média do implante dentário. A quantidade dos artefatos produzidos foi avaliada pela mensuração do desvio padrão (DP) dos tons de cinza e da taxa de contraste-ruído (CNR) das 11 regiões de interesse (ROI) que foram estabelecidas em diferentes distâncias (1,5 cm, 2,5 cm e 3,5 cm) e angulações (65º, 90º, 115º e 140º) da região do implante. Os valores de DP e CNR foram comparados por análise de variância com teste post-hoc de Tukey. Quando comparada as imagens do grupo titânio com as do grupo controle considerando cada uma das ROIs, os valores de DP não diferiram entre si em todas as regiões e os equipamentos. No entanto, ao comparar as diferentes posições das ROIs dentro do mesmo grupo de imagens, para o equipamento Picasso Trio, foram observados valores mais altos nas ROIs mais próximas ao implante (p<0,05). As imagens do grupo zircônia apresentaram os valores mais altos que os demais grupos em algumas ROIs para os equipamentos Picasso e ProMax (p<0,05). No ProMax, essa diferença foi observada mesmo nas ROIs mais distantes ao implante. Valores de CNR não foram influenciados pela ROI no Picasso, mas apresentaram valores menores nas ROIs mais próximas ao implante de zircônia para os equipamentos ProMax e Accuitomo. Foi possível concluir que a quantidade e a magnitude dos artefatos em TCFC são influenciados pelo tipo do implante e pelo equipamento de TCFC. Embora sejam mais pronunciados em regiões próximas ao implante e localizadas a 90º em relação ao longo eixo mandibular, elas podem chegar a até 3,5 m do objeto gerador de artefato

Abstract: The presence of artifacts in Cone Beam Computed Tomography (CBCT) scans is inevitable, when X-rays reach dense and high atomic number materials such as dental implants. Considering the literature consulted, there are no studies that have evaluated the magnitude of this effect for different dental implant materials. Thus, the aim in present study was to evaluate the magnitude of artifacts produced by the presence of titanium and zirconium implants in CBCT images. For this, two human mandibles were used for insertion of a titanium implant and a zirconium implant, at different times, in the region of the tooth 46. In addition, in the buccal cortical, in the region referring to the middle of the dental implant, an epoxy resin¿based tissue substitute (ERBS) block was fixed with the purpose of standardizing the selection of the images to be evaluated. Each dry human mandible was maintained in a cylindrical plastic container (16-cm diameter) by the help of an impression material. After this, the cylindrical plastic was filled with water to simulate the soft tissue radiation attenuation. The images were acquired in three CBCT devices: Picasso Trio 3D, ProMax 3D and 3D Accuitomo 80, using the same acquisition protocol: 80 kVp, 5 mA and voxel of size 0.2 mm. CBCT images were acquired before and after insertion of the implants. For the evaluation of the images, the same axial reconstruction was selected in the middle region of the dental implant, using the ERBS block as reference. The quantification of the artifacts produced was measured by the standard deviation (SD) of the gray values and the contrast-noise ratio (CNR) of the 11 regions of interest (ROI) that were established at different distances (1,5 cm, 2,5 cm and 3,5 cm) and angulations (65º, 90º, 115º and 140º) of the implant region. The SD and CNR were compared by analysis of variance with post-hoc Tukey test. Considering each of the ROIs, SD values of the images of titanium group did not differ from those of control group, in all regions and machines. However, when comparing the different positions of the ROIs within the same group of images, higher values were observed in the ROIs closer to the implant for the Picasso Trio (p <0.05). The images of the zirconium group presented higher values than the other groups in some ROIs for the Picasso and ProMax (p <0.05). In ProMax, this difference was observed even in the ROIs farthest from the implant. CNR values were not influenced by the ROI in Picasso, but showed lower values in the ROIs closest to the zirconium implant for ProMax and Accuitomo. In conclusion, the number and magnitude of artifacts in CBCT are influenced by the type of implant and the CBCT unit. Although they are more pronounced in regions close to the implant and located 90º in relation to mandibular long axis, they can reach up to 3,5 cm from artifact generator object
Subject: Tomografia computadorizada de feixe cônico
Artefatos
Titânio
Zirconio
Language: Multilíngua
Editor: [s.n.]
Citation: FONTENELE, Rocharles Cavalcante. Magnitude de artefatos de tomografia computadorizada de feixe cônico relacionados com diferentes tipos de implantes dentários: impacto na qualidade da imagem = Magnitude of CBCT image artifacts related to dental implants: impact on image quality. 2018. 1 recurso online (50 p.). Dissertação (mestrado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Piracicaba, SP.
Date Issue: 2018