Title: Impacto dos desgastes das fresas na instalação de implantes dentários : análise de dois sistemas convencionais em um modelo ex vivo
Title Alternative: Burs wears impact on dental implant installation : analysis of two conventional systems in an ex vivo model
Author: Cardoso, Pauline Magalhães, 1989-
Advisor: Asprino, Luciana, 1974-
Abstract: Resumo: Objetivo: Avaliar a influência da reutilização de fresas de sistemas de implantes dentais, sobre o aquecimento ósseo, deformação das fresas, torque de fresagem e estabilidade primária dos implantes após osteotomias em um modelo ex vivo. Materiais e métodos: Foram realizadas 40 perfurações com dois sistemas de implantes convencionas em costelas bovinas frescas. As perfurações foram realizadas com auxilio do motor cirúrgico iChiropro. Todas as perfurações foram cronometradas e um termômetro digital foi utilizado para aferir a temperatura, antes e depois das perfurações. Todas as fresas tiveram os torques de fresagem (TF) aferidos e foram submetidas à pesagem em uma balança de precisão antes de serem utilizadas (T0), após 20 utilizações (T20) e após 40 utilizações (T40). Foram submetidas também a análise de microscopia eletrônica de varredura (MEV) no T0 e no T40. Na 1º utilização (T1), T20 e T40 implantes dentários foram instalados. O torque de inserção (TI) e a frequência de ressonância (FR) dos implantes foram aferidas. Após o T40 os implantes foram submetidos a um TI de 70 Ncm (T70Ncm). Resultados: Não houve diferença de temperatura, tempo, entre os valores das massas e do TF tanto intra quanto inter-sistemas ao longo das utilizações. Na análise do MEV observou-se deformação significativa das fresas no T40. Detectou-se diferença no TI entre T1 e T40 do sistema Neodent® (P=0.002). O TI apresentado pelo sistema Neodent® no T2 também foi diferente do TI no T20 e T40 do sistema P-I Brånemark (P=0.002). Houve diferença entre a FR no T40 com o sistema P-I Brånemark (Vestíbulo/lingual) e no T70Ncm com o sistema Neodent® (Mesio/distal) (P=0.001). Não houve correlação entre o TI e a FR. Conclusão: Houve influência da reutilização das fresas no desgaste das mesmas. Não houve influência da reutilização das fresas no torque de fresagem e no peso. Os dois métodos de avaliação da estabilidade primária diferiram nos resultados: O torque de inserção foi significantemente menor com as reutilizações das fresas e a análise da frequência de ressonância não sofreu alteração

Abstract: Objective: To evaluate the influence of the dental implant burs re-use on bone heating, drill deformation, milling torque and primary stability of the implant after osteotomies in an ex vivo model. Methods: We performed 40 osteotomies with two conventional dental implants systems in fresh bone ribs. Drilling were carried out with the aid of iChiropro surgical motor drill. All perforations were timed and, a digital thermometer was used to measure the temperature before and after the perforations. All burs had the milling torque (MT) measured and submitted to weighing on a precision balance before use (T0), after 20 uses (T20) and after 40 uses (T40). They were also subjected to electron microscope analysis (SEM) at T0 and T40. In use 1 (T1), T20 and T40 dental implants were installed. The insertion torque (IT) and the resonance frequency (RF) were measured. After T40 implants were subjected to a 70 Ncm IT (T70Ncm). Results: There was no difference in temperature, time, between the masses and the TF values both intra and inter-system along the uses. There was no difference between the observed masses and TF with both systems, in no time measure (P = 0.10). In SEM analysis we found a significant deformation of the burs in T40. Detected a difference in IT between T1 and T40 of Neodent® system (P = 0.002). TI presented by Neodent® system in T2 was also different from IT in T20 and T40 P-I Brånemark system (P = 0.002). There was a difference between the RF in T40 with the P-I Brånemark system (Vestibule / lingual) and T70Ncm with Neodent® system (Mesio / distal) (P = 0.001). There was no correlation between IT and FR. Conclusion: Burs reuse influenced on their wear. There was no influence of burs reuse in the milling torque and weight. The two methods for assessing the primary stability differed in the results: The insertion torque was significantly lower with burs reuse and the analysis of resonance frequency did not change
Subject: Osteotomia
Estresse mecânico
Implantes dentários
Editor: [s.n.]
Citation: CARDOSO, Pauline Magalhães. Impacto dos desgastes das fresas na instalação de implantes dentários: análise de dois sistemas convencionais em um modelo ex vivo. 2016. 1 recurso online (73 p.). Dissertação (mestrado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Piracicaba, SP. Disponível em: <http://www.repositorio.unicamp.br/handle/REPOSIP/330566>. Acesso em: 31 ago. 2018.
Date Issue: 2016
Appears in Collections:FOP - Tese e Dissertação

Files in This Item:
File SizeFormat 
Cardoso_PaulineMagalhaes_M.pdf29.97 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.